Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Determination of Nitrogen Oxides (NOx) in Waste Incineration Flue Gas: Methods, Standards, and Quali

Accurate determination of nitrogen oxides (NOx) in waste incineration flue gas is essential for demonstrating regulatory compliance, validating continuous emission monitoring systems (CEMS), and optimizing denitrification processes such as SNCR and SCR. This technical article reviews the formation and characteristics of NOx in waste-to-energy plants, the applicable standards such as GB 18485-2014, EN 14792, ISO 10849, and EPA Method 7E, and the sampling, analytical, and quality assurance procedures used by accredited third-party laboratories and plant operators to produce defensible NOx measurement data.

Technical Overview

Nitrogen oxides (NOx) emitted from waste incineration facilities consist predominantly of nitric oxide (NO) together with a smaller fraction of nitrogen dioxide (NO2). In the atmosphere, NO is gradually oxidized to NO2, contributing to photochemical smog, acid deposition, ozone formation, and secondary fine particulate matter. During the combustion of municipal solid waste, most NOx originates from fuel-bound nitrogen (fuel NO), with additional contributions from thermal fixation of atmospheric nitrogen at the high flame temperatures typical of grate furnaces and fluidized-bed combustors. Untreated NOx concentrations in waste incineration flue gas commonly range from approximately 200 to 500 mg/m3, which is why nearly all modern plants deploy selective non-catalytic reduction (SNCR) or selective catalytic reduction (SCR) systems to meet emission limits.

NOx emissions from waste incineration are tightly regulated worldwide. In China, GB 18485-2014 (Standard for Pollution Control on the Municipal Solid Waste Incineration) sets NOx limits of 300 mg/m3 as a 1-hour average and 250 mg/m3 as a 24-hour average, referred to 11% oxygen on a dry basis at standard conditions (273.15 K, 101.325 kPa). In the European Union, the Industrial Emissions Directive 2010/75/EU specifies a daily average NOx limit of 200 mg/m3 for waste incineration plants, also referenced to 11% O2. In the United States, municipal solid waste combustors are subject to ppm-level NOx limits under 40 CFR Part 60, with performance demonstrated using EPA reference methods. Compliance is established through both periodic manual stack testing by accredited third-party laboratories and continuously operating CEMS installed at the stack.

Measurement of NOx in this matrix is technically demanding. The flue gas contains high moisture (typically 10-25%), acid gases (HCl, SO2, HF), particulate matter, and ammonia slip from SNCR systems, all of which can bias results if sampling, conditioning, analyzer selection, and quality assurance are not properly controlled according to standards such as GB/T 16157, HJ 75/HJ 76, EN 14181, and ISO 10849.

Testing Samples

Flue Gas from Municipal Solid Waste Incinerators

Representative samples are extracted from sampling ports located on the stack or downstream ducts of grate furnaces and fluidized-bed combustors, following the siting and port requirements of GB/T 16157 and equivalent international practice. Measurement planes are selected where gas flow is well mixed, typically at least two stack diameters downstream and half a diameter upstream of flow disturbances, to ensure that the sample is representative of total NOx emission.

Flue Gas from Hazardous and Medical Waste Incinerators

Hazardous and clinical waste incineration generates a flue gas with greater compositional variability and higher concentrations of corrosive species. Sampling trains and transfer lines are constructed from corrosion-resistant materials such as PTFE, quartz, or high-alloy stainless steel, and the entire sample path is heated to prevent condensation and acid dew-point corrosion that could alter NOx concentrations.

Stack Gas at SNCR/SCR Outlets and Treatment Train Interstages

Measurements at the outlet of denitrification systems are used to verify deNOx efficiency, evaluate catalyst performance, and assess reagent injection. At these locations, particular attention is paid to ammonia interference, because NH3 slip can react or be converted within the analytical system and generate a positive NOx bias.

Samples for CEMS Verification and Comparison Testing

Parallel manual measurements are performed alongside installed continuous emission monitoring systems for relative accuracy test audits (RATA) under EPA requirements, QAL2 calibration and annual surveillance tests under EN 14181, or comparison testing under HJ 75-2017, providing independent verification of the automated NOx data used for regulatory reporting.

Testing Items

A complete NOx determination program for waste incineration flue gas typically includes the following measurement and quality control items:

  • Concentration of nitric oxide (NO)
  • Concentration of nitrogen dioxide (NO2)
  • Total NOx expressed as NO2 equivalent
  • Oxygen (O2) content for correction to the 11% reference oxygen concentration
  • Moisture content of the flue gas for dry-basis reporting
  • Flue gas temperature, static pressure, velocity, and volumetric flow rate for calculating mass emission rates
  • Instrument calibration error, zero and span drift, system response time, and NO2-to-NO converter efficiency as supporting quality assurance items

Testing Methods

Chemiluminescence Analysis (EN 14792, EPA Method 7E)

Chemiluminescence is the designated reference technique for NOx determination in many jurisdictions. The sample is drawn into a reaction chamber where NO reacts with excess ozone to form excited NO2 molecules; the light emitted upon relaxation is detected by a photomultiplier tube and is proportional to the NO concentration. Total NOx is measured by passing the sample through a converter (heated molybdenum or photolytic type, typically operating around 300-400 °C) that reduces NO2 to NO before analysis. EPA Method 7E requires a converter efficiency of at least 95%, verified by gas-phase titration, and calibration error within 2% of span using certified reference gases. The method offers sub-ppm detection limits and excellent linearity, making it the benchmark for compliance testing and CEMS auditing.

Non-Dispersive Infrared Absorption (NDIR, HJ 692)

NDIR analyzers quantify NO by its characteristic infrared absorption band near 5.3 μm, often using gas filter correlation to compensate for cross-interference. NO2 is determined either by a separate optical channel or by upstream conversion to NO. NDIR instruments are robust, widely used in fixed CEMS at waste-to-energy plants, and are the basis of the HJ 692-2014 standard method for stationary source NOx in China.

Ultraviolet Differential Optical Absorption Spectroscopy (UV-DOAS)

UV-DOAS systems measure NO and NO2 simultaneously using their structured absorption features in the ultraviolet region (approximately 195-230 nm for NO, with well-defined NO2 bands in the near-UV/visible). Because both species are measured directly without chemical conversion, DOAS avoids converter-related biases and permits hot, wet-basis measurement of the undiluted flue gas.

Portable Electrochemical Method (HJ 693)

The fixed-potential electrolysis method uses electrochemical cells in which NO diffusing to a sensing electrode generates a current proportional to concentration. Portable analyzers based on this principle, as specified in HJ 693-2014, enable rapid on-site screening, verification of abatement system performance, and supplementary compliance spot checks. Cross-sensitivities to CO, SO2, HCl, and NH3, as well as temperature and pressure effects, must be characterized and compensated, and frequent calibration with certified gases is required.

Wet Chemical Spectrophotometric Methods (HJ/T 42, HJ/T 43)

Classical manual methods capture NOx in an absorbing solution after oxidation, then quantify the resulting nitrite or nitrate colorimetrically: the ultraviolet spectrophotometric method (HJ/T 42-1999) and the Saltzman naphthylethylenediamine colorimetric method (HJ/T 43-1999). These techniques remain useful as independent reference checks, although they are labor-intensive and have largely been superseded by instrumental analyzers for routine work.

Sampling, Conditioning, and Calibration Requirements

Because NO2 is highly soluble and readily lost to condensate, the entire sampling train, including the probe, particulate filter, and heated transfer line, is maintained at 120-180 °C. Sample residence time is minimized to limit NO oxidation and NO2 adsorption. Moisture is either removed using dryers demonstrated not to scrub NO2, or the analysis is performed on a wet basis with heated systems. Calibration employs zero gas plus certified NO-in-nitrogen reference gases traceable to national metrology institutes, with the span selected at approximately 1.0 to 1.5 times the applicable emission limit. Automated measuring systems are additionally subject to the suitability, calibration, and ongoing quality control regimes of EN 14181 (QAL1, QAL2, QAL3, and annual function tests), ISO 10849, and HJ 75/HJ 76, including drift limits and relative accuracy assessments against the manual reference method.

Applications

NOx determination in waste incineration flue gas supports a wide range of engineering, regulatory, and research objectives:

  • Regulatory compliance demonstration for operating permits and statutory emission reporting
  • Commissioning and acceptance performance testing of newly built or retrofitted incineration lines
  • Verification and optimization of SNCR and SCR denitrification systems, including reagent dosing control and ammonia slip management
  • CEMS installation validation, QAL2 calibration, RATA audits, and annual surveillance testing
  • Emission inventories, environmental impact assessment verification, and cleaner production audits
  • Combustion optimization studies, such as primary air distribution and furnace temperature control for low-NOx operation
  • Academic and industrial research on nitrogen conversion chemistry in waste-to-energy processes

FAQ

Why is NOx reported at 11% oxygen and standard dry conditions?

Emission limits are expressed at a fixed reference oxygen concentration so that results from different plants and operating loads are comparable regardless of excess air dilution. For waste incineration, both Chinese and European regulations specify 11% O2 on a dry volume basis at 273.15 K and 101.325 kPa. Measured concentrations are converted using simultaneously determined oxygen and moisture data, so accurate O2 and water vapor measurements are an integral part of NOx determination.

What is the difference between portable electrochemical and chemiluminescence measurements?

Chemiluminescence analyzers, as specified in EN 14792 and EPA Method 7E, provide higher accuracy, lower detection limits, and stronger traceability, and are preferred for formal compliance testing and CEMS audits. Portable electrochemical analyzers under HJ 693 offer speed and convenience for screening, process checks, and supplementary measurements, but require careful management of cross-sensitivity and frequent calibration to achieve acceptable uncertainty.

Can NO2 be measured directly, or is a converter always needed?

Direct NO2 measurement is possible with UV-DOAS and certain multi-channel infrared or FTIR analyzers. Chemiluminescence systems measure only NO directly and therefore rely on converters whose efficiency must be demonstrated to be at least 95%. Photolytic converters are preferred over heated metal converters when ammonia or other reactive nitrogen species are present, because metal surfaces can convert these interferents to NO and bias the result high.

How does ammonia slip from SNCR affect NOx results?

Ammonia injected for NOx reduction can pass unreacted into the flue gas. In heated molybdenum converters, NH3 may be partially converted to NO, producing a positive bias of up to several percent of the NH3 concentration. Control measures include using photolytic converters, placing an NH3 scrubber where validated, documenting ammonia slip, and evaluating converter selectivity during method validation.

How often should CEMS for NOx be calibrated and audited?

Typical practice includes automatic or manual zero and span checks at least daily, calibration gas audits at defined intervals, and an annual relative accuracy test against the manual reference method, with control charts maintained under QAL3 or equivalent procedures. Chinese facilities follow the calibration, drift, and acceptance criteria of HJ 75-2017 and HJ 76-2017, while European plants apply the EN 14181 quality assurance framework.

What sample conditioning is required to prevent NO2 losses?

Heated probes, heated filters, and heated sample lines maintained above the acid and water dew points prevent condensation, since NO2 dissolves rapidly in liquid water. Where condensate removal is unavoidable, coolers and permeation dryers must be qualified to demonstrate quantitative NO2 recovery, or the measurement should be performed on a hot, wet basis with a fully heated analytical path. Leak checks and short residence times further protect sample integrity.